LAWSUITS NEWS & LEGAL INFORMATION
FDA Issues Class I Recall of DePuy LPS Diaphyseal Sleeve
Washington, DC: The Food and Drug Administration (FDA) has issued a Class I recall of the DePuy LPS Diaphyseal Sleeve. The LPS Diaphyseal Sleeve to Diaphyseal Sleeve Base taper connection may not be sufficient to accommodate potential physiologic loads that may be transferred to the junction during normal gait activities by some patients. This may result in fracture of the sleeve at the taper joint which may also lead to loss of function or loss of limb, infection, compromised soft tissue or death. The FDA has received a total of 10 reports (6 fractures and 4 reports of loosening that may or may not be attributed to the same device design issue) of incidents in which the device has malfunctioned.
The LPS Diaphyseal Sleeve is intended for use with the LPS System which is an end-stage revision knee replacement product that allows surgeons to reconstruct severe soft tissue and bony defects. The diaphyseal sleeve is intended to enhance the fit and fill of the diaphyseal femoral canal with femoral and tibial replacements.
On Janury 4, 2013, DePuy issued an Urgent Medical Device Recall informing hospitals and surgeons of the problem and to immediately stop distributing or using the recalled lots. If a medical facility has the affected product in stock, it should be returned to DePuy. The affected devices were manufactured from 2008 to July 20, 2012.
Last updated on
The LPS Diaphyseal Sleeve is intended for use with the LPS System which is an end-stage revision knee replacement product that allows surgeons to reconstruct severe soft tissue and bony defects. The diaphyseal sleeve is intended to enhance the fit and fill of the diaphyseal femoral canal with femoral and tibial replacements.
On Janury 4, 2013, DePuy issued an Urgent Medical Device Recall informing hospitals and surgeons of the problem and to immediately stop distributing or using the recalled lots. If a medical facility has the affected product in stock, it should be returned to DePuy. The affected devices were manufactured from 2008 to July 20, 2012.
Legal Help
If you or a loved one has suffered similar damages or injuries, please click the link below and your complaint will be sent to a lawyer who may evaluate your claim at no cost or obligation.Last updated on
DEPUY KNEE REPLACEMENT LEGAL ARTICLES AND INTERVIEWS
Knee Replacement Triggers Airport Metal Detectors
April 7, 2013
Compared to the pain that Joseph has endured since getting a knee replacement in 2008, setting off metal detectors at airports is only a minor irritant. READ MORE
READ MORE Defective Products Settlements and Legal News
READ MORE Drugs/Medical Settlements and Legal News
READ MORE Personal Injury Settlements and Legal News
READ MORE Drugs/Medical Settlements and Legal News
READ MORE Personal Injury Settlements and Legal News